FIT screening

Colorectal cancer (CRC) is one of the most frequently diagnosed cancers. The good news is that CRC incidence and mortality can be reduced significantly if detected early enough.

Faecal immunochemical tests (FIT) are non-invasive and can detect blood in stool invisible to the naked eye. Due to its simplicity, FIT is currently considered the best non-invasive test for CRC screening.

Invest a little time in your own health by taking the FIT to prevent or detect colon cancer early on.
For further information, please visit our ‘FIT for screening’ website www.fitscreening.eu/patients

 

 

Please select your country. This will keep you up to date with your country-specific information.

CA-101

Haemostasis testing - straight forward and compact

The CA-101 uses turbo-densitometry and combines the best of both mechanical and optical detection principles. The lamp intensity automatically adjusts to sample turbidity, which means all samples are properly analysed with constant high quality results. Results are automatically calculated for PT and fibrinogen. They can be output to an optional external printer.

Standardised procedure

The instrument incubates and mixes sample and reagent automatically and indicates when to add starting reagent. The auto-start function automatically detects when reagent is added (pipetting is done manually) and triggers mixing and clot detection. This procedure requires only little manual intervention, is easy to perform and ensures consistent, high quality results so you can focus further on standardisation in your lab.

Specifications

Specifications

Detection principle:

turbo-densitometric, opto-mechanical


No. of detector channels:

1 channel


Parameter:

PT, APTT, Fbg, TT


Throughput:

PT only: about 30 tests per hour


Test protocol:

up to 7 programmable


Sample volume:

min. 50 L (PT, APTT, TT, Fbg)


Reagent position:

1


Calibration:

up to 7 curves with up to 9 points each can be stored


Interfaces:

LCD display (user interface)
serial RSC for data output to an optional external printer


Dimensions / weights:

main unit
w x h x d [mm] / [kg]
205 x 60 x 130 / approx. 0.65


Operating voltage

24V DC / 2A. Universal adapter included for 100 – 240 V