HISCL SARS-COV-2 AG ASSAY KIT

Simple workflow, high usability

The HISCL SARS-CoV-2 Ag Assay Kit provides laboratories with a robust solution that quickly delivers reliable results and increases their testing capacity. This helps smooth the SARS-CoV-2 testing workflow and further assists local healthcare systems in their efforts against the pandemic.

More details

Excellent performance

The HISCLTM SARS-CoV-2 Antigen test offers excellent sensitivity and specificity, and superb correlation with RT-PCR ct values, so you can have confidence in the results.

Comparison of the HISCL SARS-CoV-2 Ag Assay cut-off index with RT-PCR ct values in 141 positive samples and 153 negative samples (nasopharyngeal and nasal swab samples) demonstrated 100% clinical sensitivity (ct ≤ 30) and 100% clinical specificity.

Days after the onset of symptoms

RT-PCR Cut-off

RT-PCR positive (n)

HISCL positive (n)

Sensitivity

95% CI

 0–7 days

ct ≤ 30

57

57

100%

94.9%–100%

 8–35 days

ct ≤ 30

16

16

100%

82.9%–100%

 All

ct ≤ 30*

80

80

100%

96.3%–100%

 All

All*

141

105

74.50%**

66.4%–81.4%   

* To the present date, a common understanding has evolved that patients having a viral load corresponding to ct values > 30 are unlikely to infect others.

** All samples considered positive by RT-PCR incl. samples with a ct > 30 (RT-PCR ct values 30–40)

Total (n)

RT-PCR cut-off

RT-PCR negative (n)

HISCL negative (n)

Specificity

95% CI

 294

ct > 40

153

153

100%

98.1%–100%

Requirements

The HISCL SARS-CoV-2 Ag Assay Kit is to be used on the HISCL-5000 fully automated, random-access immunochemistry analyser. The assay kit and the reagents in conjunction with the HISCL technology help clinical laboratories to cope with demanding SARS-CoV-2 testing workloads.

Reagent name Item number Package size
HISCL SARS-CoV-2 Ag Assay Kit BC29960 100 tests
HISCL SARS-CoV-2 Ag Calibrator AH270120 2 vials (1 x NC and 1 x PC)* x 1 mL (4–5 tests per vial)
HISCL SARS-CoV-2 Ag Control BK261298 4 vials (2 x level 1 and 2 x level 2) x 3 mL (12–15 tests per vial)
Extract solution AD578612 5 x 10 prefilled tubes + filter tips

*NC and PC correspond to two different levels required for the calibration of this assay to calculate the cut-off values.

Technical specifications

Test technology

chemiluminescence enzyme immunoassay (CLEIA)

Dedicated instrument

fully automated, random-access immunochemistry analyser HISCL-5000

Throughput

up to 200 tests/h on the HISCL-5000

Time to result

17 minutes to the first result on the HISCL-5000

Sample type

nasal and nasopharyngeal swabs – transport as dry swabs or swabs in transport media (e.g. UTM)

Sample volume

minimum 200 µL (= 5 drops) of the sample (extracted solution)

Measurement tubes

conical 4-mL sample cup (no barcode possible) or
standard sample tube (13–16 mm diameter, 75–100 mm height) with the option of barcode labelling

Contact us
You have a question and would like to get in touch with our experts? We look forward to receiving your enquiry.

Sysmex Europe SE

Bornbarch 1

22848 Norderstedt

Germany

+49 (40) 527 26 0

+49 (40) 527 26 100

Documents

Product documents

Safety data sheets

  • Filter by language
  • Bulgarian
  • Croatian
  • Czech
  • Danish
  • Dutch
  • English
  • Estonian
  • Finnish
  • French
  • German
  • Greek
  • Hungarian
  • Italian
  • Latvian
  • Lithuanian
  • Norwegian
  • Polish
  • Portuguese
  • Romanian
  • Russian
  • Serbian
  • Slovak
  • Slovenian
  • Spanish
  • Swedish
  • Turkish